Calibration Help

Calibration Module Help

Lot Number -
Model -

Calibration Modules - Help & Support

← View All 414 Monitors

Choose Your Calibration Module

Use the CM-18 for 414R-HP resistivity monitors. Use the CM-1K for 414C conductivity and 414TDS monitors.

Resistivity

CM-18

For 414R-HP resistivity monitors when you need verification or calibration confirmation against the 18.00 megohm reference value.

Use With 414R-HP
Expected Reading 18.00 megohm at 25°C
Adjustment R16 Potentiometer
View Procedure Download Instruction Sheet
Conductivity / TDS

CM-1K

For 414C conductivity monitors and 414TDS monitors when you need verification against the 1000 µS/cm or 666 ppm calibration reference.

Use With 414C / 414TDS
Expected Reading 1000 µS/cm or 666 ppm at 25°C
Adjustment R11 Potentiometer
View Procedure Download Instruction Sheet

CM-18 Calibration Module

For 414R-HP resistivity monitors, including 110VAC, 220VAC, 24VAC, and -520 output models.

Simulated Value
18.00 MΩ ±0.018 MΩ
Resistor Tolerance
±0.01%
Cell Constant
0.05
Traceability
NIST (MIL-PRF-55182)
Adjustment
R16 Potentiometer
Note: The 414R-HP is factory calibrated. Adjustment using the CM-18 should only be performed if verification shows the reading is out of tolerance.

How to Use the CM-18

Disconnect Power and Remove the Back Cover Disconnect power to the 414R-HP and remove the enclosure back cover by loosening the 4 captive screws.
Locate the Calibration Trimmer Locate the CAL / R16 trimmer on the lower edge of the circuit board before installing the calibration module.
Disconnect the Sensor and Install the CM-18 At TB1 terminals 7-11, disconnect the 5 CS10S cell wires and insert the CM-18 pins in the same positions with the label facing up. Secure the terminal screws.
Apply Power and Stabilize Apply power and allow the display to stabilize for several minutes. Check temperature; it should read 25°C. There is no temperature adjustment.
Verify or Adjust If the displayed reading differs from 18.00 megohms, adjust the R16 trimmer with a small screwdriver until the reading agrees. The meter continuously averages readings, so allow several minutes between very small adjustments. A full calibration typically takes 10–15 minutes.
Power Down and Restore the Sensor Wiring After calibration, disconnect power, remove the CM-18, and reinstall the CS10S cell wires at TB1 terminals 7-11.
Replace the Back Cover and Power Up Reinstall the enclosure back cover, restore power, and confirm the unit returns to normal operation.
Important: Do not use the CM-18 with conductivity or TDS monitors. The CM-18 is designed exclusively for 414R-HP resistivity monitors. For 414C and 414TDS, use the CM-1K.

CM-18 Module Verification Values

Use a calibrated multimeter set to Ohms (Ω) to verify the reliability of the CM-18 module. Annual verification is recommended.

Test Points Expected Reading Tolerance
Pin 7 → Pin 8 0.900 MΩ ± 0.1%
Pin 10 → Pin 11 13.56 kΩ ± 0.1%

CM-18 Troubleshooting

My 414R-HP does not read 18.00 megohms with the CM-18 connected

If the reading is not correct:

  • Confirm the sensor is disconnected - The CM-18 must replace the CS10S sensor. Do not connect both at the same time.
  • Allow several minutes to stabilize - The meter continuously averages readings, so changes are not instantaneous. Wait for the display to settle before evaluating.
  • Check temperature - The temperature display should read 25°C ±0.1°C (24.9–25.1°C). Temperature is not adjustable.
  • Adjust using R16 - Use the R16 calibration potentiometer to adjust to 18.00 megohms. Make very small adjustments and allow several minutes between changes.
  • Verify CM-18 Calibration Values - Use a multimeter to verify the module test points: Pin 7 to Pin 8 should read 0.900 MΩ, and Pin 10 to Pin 11 should read 13.56 kΩ.
  • Inspect Module Condition - Ensure the CM-18 pins are clean, straight, and making proper contact. Replace the module if any physical damage is observed.
"OL" Reading with CM-18 Connected

An "OL" (over limit) reading indicates an open circuit condition.

Check the following:

  • Verify installation (TB1: 7–11) - Ensure the CM-18 is fully seated on the correct sensor input terminals.
  • Inspect pin contact - Confirm all pins are straight, clean, and making solid contact.
  • Verify calibration values - Confirm the module test points read correctly: Pin 7 to Pin 8 should read 0.900 MΩ, and Pin 10 to Pin 11 should read 13.56 kΩ.

If the CM-18 is correctly installed and verified to read the proper resistance, the issue is most likely with the monitor input circuit.

For further assistance, contact: orders@hydrocheck.com.

How the CM-18 Works

  • Simulates a CS10S resistivity sensor with a cell constant of 0.05 and an 18.00 megohm water equivalent for 414R-HP verification.
  • Constructed with NIST traceable resistors at ±0.01% per MIL-PRF-55182 and factory tested to simulate 18.00 megohm water within ±0.018 megohm.
  • It is a sealed module and does not require calibration.

CM-18 FAQ

How often should I verify calibration with the CM-18?

This depends on your quality system requirements. Many facilities verify calibration quarterly or semi-annually. Consult your quality or compliance team for the interval required by your specific standards.

Does the CM-18 expire or need recalibration?

No. The CM-18 is a sealed unit and does not require calibration. If the module is physically damaged or the pins are compromised, contact orders@hydrocheck.com for a replacement.

How long does the calibration process take?

The full calibration process typically takes 10-15 minutes. Most of that time is waiting for the meter's averaging to stabilize between adjustments of R16.

What should the temperature display read?

The temperature should read 25°C ±0.1°C (24.9–25.1°C). Temperature is not adjustable.

CM-1K Calibration Module

For 414C conductivity monitors and 414TDS monitors, including 110VAC, 220VAC, 24VAC, and -520 output models.

Simulated Value
1000 µS/cm ±1 µS/cm
TDS Equivalent
666 ppm ±1 ppm
Cell Constant
1.06
Traceability
NIST Traceable
Adjustment
R11 Potentiometer
Note: The 414C and 414TDS are factory calibrated. Adjustment using the CM-1K should only be performed if verification shows the reading is out of tolerance.

How to Use the CM-1K

Disconnect Power and Remove the Back Cover Disconnect power to the 414C or 414TDS and remove the enclosure back cover by loosening the 4 captive screws.
Locate the Calibration Trimmer Locate the CAL / R11 trimmer on the lower edge of the circuit board before installing the calibration module.
Disconnect the Sensor and Install the CM-1K At TB1 terminals 7-11, disconnect the CTS1 sensor wiring and install the CM-1K in the same positions as indicated on the module label. Secure the terminal screws.
Apply Power and Stabilize Apply power and allow the reading to stabilize for several minutes. Check temperature; it should read 25°C. There is no temperature adjustment.
Verify or Adjust For the 414C, the display should read 1000 µS/cm. For the 414TDS, the display should read 666 ppm. If adjustment is needed, use the R11 trimmer to adjust to the correct target and allow several minutes between small changes. A full calibration typically takes 10–15 minutes.
Disconnect the Module and Restore the Sensor Once calibration is complete, disconnect the CM-1K and reconnect the CTS1 sensor to the 414C or 414TDS.
Replace the Back Cover and Power Up Reinstall the rear cover and power up the unit to confirm normal operation.
Important: Do not use the CM-1K with resistivity monitors. The CM-1K is designed exclusively for 414C and 414TDS monitors. For 414R-HP resistivity monitors, use the CM-18.

CM-1K Module Verification Values

Use a calibrated multimeter set to Ohms (Ω) to verify the reliability of the CM-1K module.

Test Points Expected Reading Tolerance
Pin 7 → Pin 8 1.066 kΩ ± 0.1%
Pin 10 → Pin 11 13.56 kΩ ± 0.1%

The 414C and 414TDS circuits are stable and calibration is not normally required unless needed for QC or certification requirements.

CM-1K Troubleshooting

414C / 414TDS does not read correctly with the CM-1K connected

If the reading is not correct:

  • Disconnect the sensor - The CM-1K must replace the CTS1 sensor. Do not connect both at the same time.
  • Allow the reading to stabilize - The meter continuously averages readings; allow several minutes for the value to settle.
  • Verify temperature - The display should read 25°C ±0.1°C (24.9–25.1°C). Temperature is not adjustable.
  • Adjust calibration (R11) - Use the R11 potentiometer to adjust the reading to 1000 µS/cm, or 666 ppm on 414TDS.
  • Inspect module condition - Ensure CM-1K pins are clean, straight, and making proper contact.
"OL" Reading with the CM-1K Connected

An "OL" (over limit) reading indicates an open circuit condition.

Check the following:

  • Verify installation (TB1: 7–11) - Ensure the CM-1K is properly installed on the correct sensor input terminals.
  • Inspect pin contact - Confirm all pins are straight, clean, and making solid contact.
  • Verify calibration module resistance values - Use a multimeter to verify the module test points: Pin 7 to Pin 8 should read 1.066 kΩ, and Pin 10 to Pin 11 should read 13.56 kΩ.

If the CM-1K is correctly installed and verified to read the proper value, the issue is most likely with the monitor input circuit.

For further assistance, contact: orders@hydrocheck.com.

How the CM-1K Works

  • Simulates a CTS1 conductivity sensor with a cell constant of 1.06, displayed as 1000 µS/cm on 414C and 666 ppm on 414TDS.
  • Constructed with NIST traceable resistors at ±0.01% and factory tested to simulate 1000 µS/cm or 666 ppm at 25°C.
  • It is a sealed module and does not require calibration.

CM-1K FAQ

Can I use the CM-1K for both the 414C and 414TDS?

Yes. The CM-1K works with both models. The 414C should read 1000 µS/cm and the 414TDS should read 666 ppm.

Does the CM-1K expire or need recalibration?

No. The CM-1K is a sealed unit and does not require calibration. If the module is physically damaged, contact orders@hydrocheck.com for a replacement.

Why does the 414TDS read 666 ppm instead of 1000?

The CM-1K simulates a 1000 µS conductivity signal. The 414TDS converts conductivity to TDS using a standard conversion factor, which yields approximately 666 ppm.

What should the temperature display read?

The temperature should read 25°C ±0.1°C (24.9–25.1°C). Temperature is not adjustable.

Need help? Our support team is ready to assist with calibration questions or replacement parts. Have your model and lot number ready if available.
orders@hydrocheck.com
Certificate of Conformance
Lot Number -
Model CM

General Certificates

Download standard Certificates of Conformance for calibration modules.

CM-18 Calibration Module

Standard compliance documentation for the resistivity calibration module.

CM-1K Calibration Module

Standard compliance documentation for the conductivity or TDS calibration module.

Traceability Lookup

Enter the lot or serial number exactly as printed on the module label to retrieve traceability and calibration data.

CM-18 Calibration Module

Format guide: CM18-xx-yymmdd, CM18-xxx-yymmdd, or CM18-Bxx-yymmdd.

Contact Us

Contact Us
First
Last

For complete product information, visit the 414 Monitor Help page or contact orders@hydrocheck.com.

1996 - 2026

Celebrating 30 Years of Cleaner Water

© 2026 Hydro-Check Systems, Inc. All Rights Reserved.

Copyright © 2026 Professional Water Quality Management Instruments, All Rights Reserved.

Return/Refund Policy