414 Resistivity Help

414 Resistivity Help

Lot Number -
Model -

414R-HP Resistivity Monitor - Help & Support

← View All 414 Monitors
Range
0.015 - 18.18 MΩ·cm
Accuracy
±1%
Sensor
CS10S
Temp Comp
Auto (5-50°C, referenced to 25°C)
Calibration
CM-18

What Does My Display Mean?

Green = Above Setpoint (Normal Operation) The monitor is indicating that the measured resistivity is above the user-defined setpoint. Water purity is acceptable and the relay is de-energized.
Red = Below Setpoint (Alert Condition) The resistivity has dropped below the user-defined setpoint. The relay will energize after the configured time delay.
OL
"OL" - Out of Range The monitor will display "OL" when the reading exceeds the upper measurable range or if the sensor is disconnected. Note: the 414R-HP will always display OL when no sensor is connected, which is normal behavior.
-UL
"-UL" - Under Limit The resistivity is below the measurable range of the instrument. This may indicate very low purity water or a sensor issue.

Status is based on the user-defined setpoint, not the full measurement range. Note: higher resistivity indicates higher water purity.

Troubleshooting

My 414R-HP is showing "OL" - what does that mean?

"OL" on the 414R-HP means the reading is Out of Range. Unlike the 414C and 414TDS, the 414R-HP will always show OL when no sensor is connected, which is normal behavior. Follow these troubleshooting steps:

  • Confirm the water is not exceeding 18.18 MΩ·cm - This is the upper limit of the 414R-HP measurement range. If water purity exceeds this value, OL is expected behavior.
  • Check for air bubbles or poor sensor contact - Ensure the sensor is fully submerged in constant-flowing water. Tap the probe gently to dislodge air. Install horizontally when possible.
  • Verify water temperature is within range - Water must be within 5-50°C for accurate measurement.
  • Verify you're using the correct sensor - The 414R-HP uses the CS10S (stainless steel body), not the CTS1 (plastic body). Using the wrong sensor will cause errors.
  • Test the CS10S sensor with a multimeter - Check sensor continuity and resistance per expected values. Refer to factory specs or contact support if unsure.
  • Check for a component failure - If OL appears with a known-good sensor and flowing water, the monitor may need repair. Contact support.
My 414R-HP is showing "-UL" - what does that mean?

"-UL" means the resistivity reading is below the measurable range of the instrument, below 0.015 MΩ·cm. This typically indicates very low purity water.

  • Verify water quality - If the water source is expected to have low resistivity, high conductivity, "-UL" may be normal. A conductivity monitor like the 414C may be more appropriate for that application.
  • Confirm you are using the correct sensor - The 414R-HP requires the Hydro-Check labeled CS10S resistivity sensor. Using the wrong sensor will produce incorrect readings.
  • Check sensor wiring at TB1 (terminals 7-11) - Verify all connections are secure and match the wiring diagram in the user manual.
  • Inspect the sensor for damage or contamination - A fouled, damaged, or improperly installed sensor can produce under-range readings. Clean or replace as needed.
  • Contact support - If "-UL" persists with known-good water and a verified sensor, contact orders@hydrocheck.com for further assistance.
Need help? Our support team is ready to assist with installation, troubleshooting, or replacement parts. Have your model and lot number ready if available.
orders@hydrocheck.com

Manuals & Instructions

Download User Manual (PDF) Data Sheet (PDF) CM-18 Calibration Instructions (PDF)

Complete installation, operating, and calibration instructions for the 414R-HP Resistivity Monitor.

Installation

Mount the Monitor The 414R-HP may be panel-mounted or wall-mounted using the optional WMA Wall Mount Adapter. Install in a dry, accessible location near the process line.
Install the CS10S Sensor Thread the CS10S sensor into the process line using the included fitting. Install horizontally when possible to minimize air pockets. Ensure the sensor is fully submerged in constant-flowing water.
Wire the Sensor to the Monitor Connect the CS10S sensor wires to terminal block TB1 (terminals 7-11) following the wiring diagram in the user manual.
Connect Power Configure the unit for 110VAC or 220VAC as required and wire input power to TB1 per the user manual.
Bleed the Line Run water through the system to remove air from the line. Allow the system to stabilize before relying on readings.
Important: Only Hydro-Check labeled CS10S sensors are designed for use with the 414R-HP. Other cells will not perform correctly and may produce inaccurate or erratic readings.

For detailed wiring diagrams and terminal assignments, see the User Manual (PDF).

How It Works

  • The CS10S sensor is installed inline in your water process and measures the electrical resistivity of the water.
  • The 414R-HP continuously displays the resistivity value in megohm-centimeters (MΩ·cm) on the front panel.
  • The instrument uses automatic dual-range measurement to maintain optimal resolution and accuracy across the full scale.
  • A linear thermistor network provides automatic temperature compensation to 25°C across a range of 5-50°C, with ±0.1°C temperature accuracy.
  • The relay changes state based on the user-defined setpoint: above setpoint = relay is de-energized; below setpoint = relay is energized. A configurable 0-120 second time delay prevents false triggers during startup or rinse-up.
  • The relay can drive a 414-RMA Remote Alarm Module, solenoid valve, BMS input, or other external equipment.

Frequently Asked Questions

What is the measurement range of the 414R-HP?

The 414R-HP measures resistivity from 0.015 to 18.18 MΩ·cm, covering applications from lower purity feedwater through high purity and ultra-pure water systems.

The instrument uses automatic dual-range measurement to maintain optimal resolution and accuracy across the full scale. With a theoretical maximum of 18.2 MΩ·cm at 25°C, the 414R-HP is calibrated to 18.18 MΩ·cm within ±1% accuracy.

Does the 414R-HP have temperature compensation?

Yes. The 414R-HP uses a linear thermistor network for automatic temperature compensation to 25°C across a range of 5-50°C. Resistivity accuracy is ±1% of reading, and temperature accuracy is ±0.1°C.

Which sensor does the 414R-HP use?

The 414R-HP uses the CS10S sensor with a stainless steel probe body. This is different from the CTS1 sensor used by the 414C and 414TDS. Make sure you're using the correct sensor, as the cell constant is different between the CS10S and CTS1.

Available in standard cable lengths (10', 25', 50', 100'). Custom lengths may be available upon request.

What accessories do I need for a complete 414R-HP setup?

Minimum required:

  • 414R-HP Resistivity Monitor
  • CS10S Resistivity Sensor (Hydro-Check labeled)
  • PC-110 Power Cable

Optional:

  • WMA Wall Mount Adapter
  • CM-18 Calibration Module
  • 414-RMA Remote Alarm (includes RPC cable)
How do I calibrate the 414R-HP?

The 414R-HP is factory calibrated and typically does not require calibration. If verification or compliance calibration is needed, use the CM-18 Calibration Module.

Connect the CM-18 in place of the sensor and refer to the CM-18 instructions included with the module. Calibration modules come with a Certificate of Conformance.

Can I connect a remote alarm to the 414R-HP?

Yes. The 414R-HP has a built-in SPDT relay on terminal block TB1 (terminals 4-6) that can drive a 414-RMA Remote Alarm Module. The RMA provides visual (green/red LED) and audible alerts at a remote location.

You'll need the 414-RMA module and an RPC cable (available in various lengths).

Documentation

User Manual PDF Data Sheet - 414R-HP Resistivity Monitor PDF CM-18 Calibration Instructions PDF
Certificate of Conformance
Lot Number - | Model: 414R-HP

General Certificate

Download a standard Certificate of Conformance for 414 Resistivity Monitors.

Lot-Specific Certificate

Certificate with traceability data for your specific lot number.

Scan your QR code or contact orders@hydrocheck.com for lot-specific certificates.

Contact Us

Contact Us
First
Last

For complete product information and specifications, visit the 414R-HP product page.

1996 - 2026

Celebrating 30 Years of Cleaner Water

© 2026 Hydro-Check Systems, Inc. All Rights Reserved.

Return/Refund Policy